6. Close the setup screen and go to the terminal.
Note: Make sure that only one instrument is connected during the setup.
7. Make sure that the correct COM port is selected.
8. Click on "Send Lead Command". Do not confirm with OK.
9. Set the instrument power to off. Wait 2 seconds and set the instrument to on again.
10. Confirm "Load Delay: 2 Sec." within 30 seconds.
11. When the "load" command was sent successfully, the menu that follows shows: ---MAIN MENU---
RMCA [2082375-1E]
12. Type the number 1 to 9 and do the prompts at the command line that follow to select a specific
operating parameter. Refer to
1. UNIT ID32 [0...32
2. COUNT PERIOD00:10 [MIN,SEC]
3. CAL LIMITS0800–1200 [LOWER, UPPER mV]
4. COUNT MODEMANUAL [AUTO, MANUAL]
5. PANEL DISPLAY2.0 [CUMULATIVE 9U0]
6. FLOW RATE 100 mL/min]
7. CALIBRATE
8. DEFAULT MEMORY
9. SETUP ANALOG I/OQ = QUIT
13. Select "9" to configure the analog inputs and outputs. Select "1" to set the analog inputs to on or
off. Push "2" to go out of the output configuration menu.
14. Select "1" for analog inputs or "2" for analog outputs, then enter the command number but do not
push the <Enter> key.
• ENTER ANALOG OUTPUT CHANNEL [0 to 7]
• ENTER LOWER SIZE
• ENTER UPPER SIZE (0=cumulative)
• ENTER FULL SCALE COUNT (0=disable channel)
Note: Enter the lowest particle size and 0 for no upper limit. The upper size specifies the count value shown by
20 mA.
15. Complete the sequence with "Q <Enter>". All of the configuration information is saved in the
instrument memory.
Note: To go out of the configuration menu at the very end, stop the sequence with "P <Enter>".
16. Do steps
11
to
15
17. Do steps
2
to
11
again for more instruments.
Menu number Command line message
1
ENTER ID
2
ENTER MIN:SEC
3
ENTER LOWER THRESHOLD
(mV)
ENTER UPPER THRESHOLD
(mV)
4
ENTER COUNT MODE (A OR
M)
2
The default ID number is "32". This number is set at the factory to a lower number. "32" typically
shows that the EPROM was changed. If the instrument shows "32", contact technical support.
18 English
Table
2.
2
]
again for more channels.
Table 2 Operating parameter selection
Description
Enter the Modbus address.
Set count period when in "AUTO" mode.
Calibration check threshold. Do not change the value.
Calibration check threshold. Do not change the value.
Auto = locally self-timed; Manual = count period is controlled
by data collection software.