8. Install the nut and then the compression fitting to the other end of 18-in. length of the black
flexible tubing. Refer to
9. Connect the compression fitting to the water weir outlet. Refer to
10. Connect the quick-connect fitting to the sensor inlet port of the instrument. Refer to
items 2 and 12.
11. Attach the quick-connect fitting to a 12-in. length of ¼-in. black flexible tubing. Refer to
items 2 and 3.
12. Attach the quick-connect fitting to the sensor outlet port on the instrument. Refer to
items 1 and 2.
13. Put the other end of the 12-in. length of ¼-in. black flexible tubing into the hole of the adjustment
cap. Make sure that the end of the tubing is 6.4 mm (0.25 in.) inside the cap. Refer to
item 4.
14. Install the drain line on the water weir. Install a clear ½-in. I.D. hose over the barbed fitting on the
water weir drain. Refer to
15. Use the drain hose to measure the distance between the water weir drain and the nearest waste
drain. Remove the excess length of the drain hose. Install the other end of the hose on the drain.
Refer to
Figure
5.
16. Open the shut-off valve on the plumbing tap and examine the tubing for leaks.
10 English