Specifications

Specifications are subject to change without notice.
Specification
Dimensions (W x D x H)
Weight
Operating temperature
Storage temperature
Enclosure
Fittings
Power requirements
Maximum solution pressure
Flow rate
Particle size
Indicators
Distance from computer to sensor
Flow control (optional)
Digital communication
Hach UDG1000 software (optional) Organize instrument readings on a SCADA system
Analog input/output card (optional)
Pollution degree
Installation category
Protection class
Operation
Troubleshooting
Details
350 x 211 x 178 mm (13.8 x 8.3 x 7.0 in.)
4.89 kg (10.78 Ibs)
0 to 50 °C (32 to 122 °F); maximum relative humidity 80% for temperatures
with a maximum of 31 ºC (87.8 °F) decreasing linearly to 50% relative
humidity at 40 ºC (104 °F)
0 to 50 °C (32 to 122 °F)
NEMA 4X (indoor use)
Quick-connect. Connect to ¼-inch O.D tubing
115-240 VAC, 50/60 Hz, 1A
65 psig, not more than 1 minute duration; 55 psig continuous
100 mL/minute nominal
Smallest size: 2 µm
Largest size: 750 µm
Power, counting display, cleaning sensor and alarm
1219.20 m (4000 ft.) maximum, entire RS485 signal path
Passive control devices available to control water flow through the
instrument
Modbus ASCII; WQS Vista data collection software (optional): Monitor filter
performance and make reports. Refer to the software documentation for
computer requirements.
Input: accepts input signals from external devices
Output: supplies an analog output level proportional to total number of
particles counted (raw count)
2
I
III (supplied by SELV power source)
on page 17
on page 24