Fora Diamond GD50 Manuel De L'utilisateur page 13

Publicité

• AC (
) - no food intake for at least 8 hours.
• PC (
) - 2 hours after a meal.
• QC (
) - testing with control solution.
3. Obtain a blood sample. (m)
Use the pre-set lancing device to puncture your desired
site. After penetration, discard the first drop of blood with a
clean tissue or cotton. Gently squeeze the punctured area to
obtain another drop of blood. Be careful NOT to smear the
blood sample. The volume of blood sample must be at least
0.5 microliter (μL) for blood glucose testing.
4. Apply the blood sample. (p)
Move your finger to meet the absorbent hole of the test
strip and the drop will be automatically be drawn onto the
test strip. Remove your finger until the confirmation window
is filled. The meter begins to count down. Do not remove
your finger until you hear a beep sound.
5. Read your result. (q)
The results of your blood glucose test will appear after
the meter counts down to 0. The results will be stored
automatically in the meter memory.
Announcement of Test Result by Universal
Tone
The blood glucose results will be broken down into individual
digits and each digit represents the corresponding number
of beeps.
The result is announced three times in succession and each
time is preceded by two quick beeps. So you will hear: 2 quick
beeps – results – 2 quick beeps – results – 2 quick beeps –
result.
For mg/dL meters, the hundreds are always announced, even
when the result is below 100.
Examples:
80 mg/dL is announced as 1 long beep (0) – 1 single pause – 8
single beeps (8) – 1 single pause – 1 long beep (0)
182 mg/dL is announced as 1 single beep (1) – 1 single pause
– 8 single beeps (8) – 1 single pause – 2 single beeps (2)
For mmol/L meters, the tens are always announced, even
when the result is below 10. The decimal point is represented
by 1 quick beep.
Examples:
6.0 mmol/L is announced as 1 long beep (0) – 1 single pause –
6 single beeps (6) – 1 single pause – 1 quick beep (.) – 1 single
pause – 1 long beep (0)
Note:
Information or warnings in the form of symbols displayed
together with the results are not announced acoustically.
Disposing Used Test Strip and Lancet
To remove the used test strip, simply push the Test Strip
Ejector button upward to eject the used test strip. The device
will automatically turn off after the test strip is removed.
EN-8