Before Testing
Control Solution Testing
Our Control Solution contains a known amount of glucose
that reacts with test strips and is used to ensure your device
and test strips are working together correctly.
Test strips, control solutions, or sterile lancets may not be
included in the kit (please check the contents on your product
box). They can be purchased separately. Please make sure you
have those items needed for a blood glucose test beforehand.
Do a control solution test when:
you first receive the device.
9
9
you begin using a new vial of test strips.
9
9
you suspect the device or test strips are not working
9
9
properly.
your blood glucose test results are not consistent with how
9
9
you feel, or if you think the results are not accurate.
you have dropped or think you may have damaged the
9
9
device.
To perform the control solution test, do the following:
1. Insert the test strip into the test slot of the device. Wait for
the device to display the test strip "
" ". (e)
2. Press MAIN to mark this test as a control solution test. With
"QC" displayed, the device will store your test result in the
memory under "QC". If you press MAIN again, the "QC" will
disappear and this test is no longer a control solution test. (f)
3. Apply the control solution. Shake the control solution vial
Note:
To avoid contaminating the control solution, do not directly
apply the control solution onto a strip.
4. Read and compare the result. After counting down to 0, the
" and blood drop
Note:
• Do NOT test your blood.
• The control solution range printed on the test strip vial or
EN-5
Important!
When doing the control solution test, you have to mark it so
that the test result will NOT mix with the blood glucose
TEST RESULTS stored in the memory. Failure to do so will
mix up the blood glucose test results with the control
solution test results in the memory.
thoroughly before use. Squeeze out a drop and wipe it off,
then squeeze another drop and place it on the tip of the vial
cap. Hold the device to move the absorbent hole of the test
strip to touch the drop. Once the confirmation window is
filled completely, the device will begin counting down. (g)
test result of the control solution will appear on the display.
Compare this result with the range printed on the test strip
vial or individual foil pack and it should fall within this range.
If the test result is out of range, read the instructions again
and repeat the control solution test. (h)
individual foil pack is for control solution use only. It is not a
recommended range for your blood glucose level.