• Install the analyzer in a clean, dry, well-ventilated and temperature-controlled location.
• Install the analyzer as near to the sampling point as possible.
• Do not install the analyzer in direct sunlight or near a heat source.
• Make sure that there is sufficient clearance to make plumbing and electrical connections.
• Make sure to leave sufficient space in front of the analyzer to open the analyzer door.
Refer to the expanded user manual on the manufacturer's website for additional information.
• Make sure that the ambient conditions are within operating specifications. Refer to
on page 3.
Although the analyzer is not designed for use with flammable samples, some EZ-analyzers use
flammable reagents. If the analyzer uses flammable reagents, make sure to obey the safety
precautions that follow:
• Keep the analyzer away from heat, sparks and open flame.
• Do not eat, drink or smoke near the analyzer.
• Use a local exhaust ventilation system.
• Use spark and explosion-proof appliances and lighting system.
• Prevent electrostatic discharges. Refer to
on page 13.
• Fully clean and dry the instrument before use.
• Wash hands before breaks and at the end of the working period.
• Remove contaminated clothing. Wash clothing before reuse.
• These fluids must be handled in accordance with local regulatory agency requirements on
permissible exposure limits.
5.2 Mechanical installation
5.2.1 Attach the instrument to a wall
Personal injury hazard. Make sure that the wall mounting is able to hold 4 times the weight of the
equipment.
Personal injury hazard. Instruments or components are heavy. Use assistance to install or move.
Personal injury hazard. The object is heavy. Make sure that the instrument is securely attached to a
wall, table or floor for a safe operation.
Attach the instrument upright and level on a flat, vertical wall surface. Install the instrument in a
location and position where the user can easily disconnect the instrument from the power source.
Refer to the illustrated steps that follow. Mounting hardware is supplied by the user. Make sure that
the fastening has sufficient load bearing capacity (approximately 160 kg, 353 lb). The wall plugs must
be selected and approved to suit the properties of the wall.