Télécharger Imprimer la page

Kartsana SILVER+ Serie Manuel D'utilisation page 32

Publicité

Instruction manual
SILVER+ R-1200
3.3 Electrical connection
The electronics are ready to support a 12V tension.
The Silver+ rail must be connected to the vehicle's electrical system with an
electrical cable of at least 4 mm
or - poles). The purpose of this connection is for the Silver+ stretcher to charge
the batteries when inside the ambulance.
It is recommended that this cable be protected on the outside with a plastic sheath to avoid
electrical problems (chafing, short circuits, etc.).
The electrical system has to be secured by means of a 20A fuse. (Built-in behind the
central rear cover).
32
cross-section (pay special attention to the +
2
It must be ensured that the
is as indicated in the figure, otherwise the
stretcher will not be loaded into the vehicle.
positive pole
REV 1
05-24

Publicité

loading

Ce manuel est également adapté pour:

Silver+ r-1200