Instruction manual
SILVER+ R-1200
3. MOUNTING THE RAIL ON THE AMBULANCE
3.1 Vehicle configuration
These installation instructions are indicated for the attachment of the SILVER+ (TG-1200)
set with the Kartsana SILVER+ (R-1200) rail.
The correct use and operation of the SILVER+ (TG-1200) stretcher depends on the correct
installation of the fastening system (rail R-1200).
The fixed part of the rail shall be fixed at a maximum distance of 50 mm
door of the ambulance.
R-1200
SILVER+ Rail
Careful consideration should be given to the distance/type of bumper that the ambulance
model is fitted with, so that if the bumper protrudes more than 180 mm it will have to be
adapted to allow the stretcher legs to be raised correctly.
Depending on the ambulance model and the type of rear bumper, the fixing dimensions
may vary and/or be adapted.
The height of the ambulance floor in relation to the ground must not exceed
70 cm.
30
A
Max.
70cm
(A)
from the rear
TG-1200
SILVER+ Stretcher
REV 1
05-24