SPEED TRANSMISSION & ADRESS SELECTION
An "ENTER" in the previous step makes two numbers appear on the display
separated by a hyphen with the first one flashing.
The number on the left (1 digit) corresponds to the baud rate:
[1 = 1200, 2 = 2400, 3 = 4800, 4 = 9600 and 5 = 19200 baud. ]
The number on the right (2 digits) corresponds to the address of the instru-
ment, which can be programmed between 00 and 99.
Successively press the key
press
to proceed to programming the next digit.
Carry out these operations until the desired parameters are displayed on the
display and press
to save the programmed data in memory and return to
ENTER
work mode.
NOTE: If address 00 is programmed, the instrument only accepts commands that do not return data or change para-
meters. For example the TARE.
PROTOCOL SELECTION
Press "ENTER" to access the programming menus and twice
instrument in the phase represented in the figure
(indication 5 trAnS, leds B and PROG lit). In this menu the communication
mode between the instrument and the D.T.E. will be selected. (Data Terminal
Equipment)
ENTER
Press
to enter this menu, or
: to skip from this menu and return to the CnF menu.
ESC
: to return to work mode.
A flashing number appears on the display (1,2 or 3 depending on the previous
selection) corresponding to the current communication protocol
[ 1 = ASCII protocol, 2 = ISO 1745 protocol and 3 = MODBUS ].
If the value on the display corresponds to the desired option, press
ESC
ENTER
or
to return to work mode.
Otherwise, press the key
: to change the number and
ENTER
: to save the data in memory and go to work mode.
KOSMOS SERIE
to vary the value of the flashing digit and
DISEÑOS Y TECNOLOGÍA, S.A.
Xarol, 6-B
P.I. Les Guixeres
08915 Badalona (Barcelona) - Spain.
www.ditel.es
PROG
A
B
RU
TARE
N
INP1
PRG
TARE
TEACH
to place the
Tel. +34 933 394 758
Fax +34 934 903 145
Email: comercial@ditel.es ; web: www.ditel.es
RS2
1
2
3
HOLD
LIMIT
MAX
MIN
DATA
4
DSP1
INP2
DSP2
FLT
STORE
RESET
LIMIT
MAX/MI
ENTER
N
DATA
ESC
31