To navigate to the menu or to configure the settings:
1. Press the
or
button.
2. Press the
button to confirm the function/input chosen.
To exit the menu:
● Choose the menu item Back in the sub-menus.
● Choose Measuring in the main menu.
4.8
Menu items
4.8.1 Measurement results
Shows all saved results with measurement unit, date and
time, as well as sample number and the information as to
whether the sample was taken manually or passively.
4.8.2 Infoscreen
Shows the part number and the serial number of the device,
as well as the serial number of the EC sensor.
4.8.3 Next service
Displays the time until adjustment or service is due in days.
4.8.4 Date and time
Sets the date and time.
1. Change the digits using the
2. Confirm with the
button.
4.8.5 Contrast
Adjusts the contrast.
1. Change the contrast setting using the
2. Confirm with the
button.
4.8.6 Loudspeaker
Adjusts the volume.
1. Set the volume using the
2. Confirm with the
button.
4.8.7 Language
Sets the language.
1. Select the language using the
2. Confirm with the
button.
4.8.8 Measurement unit
Sets the measurement unit.
1. Select the measurement unit using the
2. Confirm with the
button.
4.8.9 Software version
Displays the version number and part number of the software.
4.8.10 Signing in
For service personnel only.
1. Enter the four-digit code to open the service menu.
®
4.8.11 Bluetooth
(optional)
®
Turns the Bluetooth
feature on or off.
®
1. Confirm Bluetooth
On or Off with the
Instructions for use
|
Dräger Alcotest 6000
or
button.
or
button.
or
button.
or
button.
or
button.
button.
Printing the measurement results
The device is prepared for printing and connecting to an
app via Bluetooth®.
®
An active Bluetooth
corresponding symbol on the display.
®
Bluetooth
must not be activated in aeroplanes.
4.8.12 Turning off
Turns off the device.
The device will turn itself off 4 minutes after last being
operated.
1. Briefly press the
button.
✓ The device turns off.
5
Printing the measurement results
®
Bluetooth
can be optionally configured on the device. To
print out the measurement results, use Bluetooth
the device to the Dräger Mobile Printer Bluetooth
number: 3700421).
Observe the instructions for use for the Dräger Mobile
®
Printer Bluetooth
.
®
Bluetooth
must not be activated in aeroplanes.
6
Troubleshooting
6.1
Troubleshooting
In the event of a device malfunction, an error message will be
shown. The number which appears underneath the message
is used for service functions. If the malfunction does not
disappear after turning the device on and off several times,
contact DrägerService.
Malfunc-
Cause
tion/Error
Device cannot be
The battery is flat.
turned on.
Volume too
The test person is
small/sample
blowing unevenly
invalid
or not strongly
enough.
Device outside
The device is too
temperature
hot or too cold.
range
The measuring
(measuring range
range has been
exceeded)
exceeded. Resid-
ual mouth alcohol
can lead to incor-
rect measured val-
ues.
Sample interrup-
Irregular exhala-
tion
tion, abrupt ending
or inhalation
connection is displayed as a
®
to connect
®
(part
Remedy
Change the battery.
Blow into the funnel
with more force and
without interruption.
Allow the device to
cool down or warm
up.
At least 15 minutes
wait time must have
passed since alcohol
was last consumed
through the mouth.
To repeat, press the
button.
Repeat the breath
sample by pressing
the
button
|
en
13