OSYPKA PACE 203 H Manuel D'utilisation page 127

Table des Matières

Publicité

Note 1:
At 80 MHz and 800 MHz, the safety distance for the higher frequency range applies.
Note 2:
The ISM bands (for industrial, scientific and medical applications) between 150 kHz and
80 MHz are 6.765 MHz to 6,795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3:
The compliance level in the ISM frequency bands between 150 kHz and 80 MHz and in
the frequency range 80 MHz to 2.5 GHz is intended to reduce the likelihood that
mobile/portable communications devices cause interference if they are unintentionally
brought into patient areas. For this reason greater safety distance is recommended
separation distance in these frequency ranges (factor 12 instead of 0.35).
Note 4:
These guidelines do not necessarily apply in all situations. The propagation of
electromagnetic waves is influenced by absorption and reflection from structures, objects
and humans.
Table 69: Safety distances between portable and mobile RF communication equipment
Page 127 of 136
PACE 203 H