PACE 203 H
Parameter
Impédance d'entrée:
Impédance de sortie:
Charge plage d'impédance:
Délais A-V
(AVD):
Délais P-V
(PVD):
Intervalle:
intervalle d'échappement
auriculaire
(Délais V-A):
intervalle auriculaire
prolongée d'échappement
(V-A Retard après la première
PVC):
Période réfractaire atriale
(ARP):
Période réfractaire
ventriculaire
(PRV):
PRAPV:
partie absolue:
partie relative:
PACE 203H V2.0
22 k ± 20 %
< 10
200 ...2000
5 ms ... 400 ms ± 5 % ± 2 ms
(minimum; 30 ms when atrial Auto Sense is
activated)
Determined by A-V Delay:
PVD = A-V Delay - 30 ms ± 5 % ± 2 ms
(minimum 5 ms when atrial Auto Sense is
not activated; minimum 30 ms when atrial
Auto Sense is activated);
may be prolonged depending on MTR
Determined by Basic Rate ± 5 %
Determined by Basic Rate and A-V
interval ± 5 %
Determined by Basic Rate and A-V Delay
(minimum 850 ms) ± 5 %
250 ms ... 400 ms ± 5 % (AAI, AAT);
A-V interval plus PVARP (DDD, VDD, DAI,
VAT, DAT)
250 ms ... 500 ms ± 5 %
(Firmware version 1.24 and up)
250 ms ± 5 % (otherwise)
100 ms ...500 ms ± 5 %
90 ms ± 5 %
PVARP - 90 ms ± 5 %
500 ms ± 5 % (or maximum possible period)
85 ms ± 5 % + 2 ms / -6 ms
after atrial as well as ventricular stimulation
and sensing
85 ms ± 5 % + 2 ms / - 6 ms
after ventricular stimulation and sensing,
55 ms ± 5 % + 2 ms / - 6 ms
after atrial stimulation
45 ms ± 5 % ± 2 ms
40 ms
Page 114 of 136