Therapy pressure: 4 to 30 cmH
Resistance with Anti-Asphyxia Valve closed to atmosphere
Drop in pressure measured (nominal) at 50 L/min: 0.2 cmH
Drop in pressure measured (nominal) at 100 L/min: 1.0 cmH
Resistance may vary due to the flexible mask frame design.
Resistance with Anti-Asphyxia Valve open to atmosphere
Inspiration at 50 L/min: 0.2 cmH
Expiration at 50 L/min: 0.4 cmH
Anti-Asphyxia Valve open-to-atmosphere pressure: <4 cmH
Anti-Asphyxia Valve closed-to-atmosphere pressure: <4 cmH
Sound: Declared dual-number noise emission values in accordance with ISO4871:1996
and ISO3744:2010. A-weighted sound power level is 25 dBA, with uncertainty of 3 dBA. A-
weighted pressure level at a distance of 1 m is 18 dBA, with uncertainly of 3 dBA.
Environmental conditions
Operating temperature: 41°F to 104°F (5°C to 40°C)
Operating humidity: 15% to 95% RH non-condensing
Storage and transport temperature: -4°F to 140°F (-20°C to +60°C)
Storage and transport humidity: up to 95% RH non-condensing
Service life: The service life of the mask system is dependent on the intensity of usage,
maintenance, and environmental conditions to which the mask is used or stored. As this
mask system and its components are modular in nature, it is recommended that the user
maintain and inspect it on a regular basis, and replace the mask system or any
components if deemed necessary or according to the instructions in the 'Cleaning your
mask' section of this guide.
Storage
Ensure that the mask is thoroughly clean and dry before storing it for any length of time.
Store the mask in a dry place out of direct sunlight.
Disposal
This product and packaging do not contain any hazardous substances and may be disposed
of with your normal household refuse.
Reprocessing the mask between patients
Only Sleep Lab Mask (SLM) variants of the mask system are intended for multi-patient re-
use. When using between patients, these masks must be reprocessed according to
instructions available on ResMed.com/downloads/masks.