The set-up place must be level and horizontal.
For versions with legs, use the adjustable legs to make sure that the product stands
level and upright, see Fig.2.
For versions with castors, the locking devices of the two front castors must be activated,
when the product is in place, see Fig.2. The base must be level, and the product may not
be placed on frames or the like.
Fig. 2
It is ideal to install the product as close as possible up against the wall, however max. 75
mm from the wall.
There must be sufficient ventilation and free air circulation around the cabinet, as illustrated
on Fig.3. There must be a minimum clearance of 250 mm above the cabinet, and 20 mm at
the sides.