5.7.3
Peak value function
This function displays the highest load value (peak value) of a weighing.
Measuring frequency: 200ms
Attention:
Never exceed the maximum permitted load of the peak value on the
balance (!!Danger of breaking!!).
• Switch on balance using the ON/OFF/TARE–button
• Press the HOLD-button and keep it pressed until „H x" in the display appears
• Press the ON/OFF/TARE until „H6" in the display appears
• Save the setting by shortly pressing the HOLD-button
• Suspend a load
• The peak value function is displayed by a [] upside left
A negative value [-] is displayed as soon as the balance is relieved
6. Menu
When the balance is switched off, press the HOLD-key and keep it pressed
Do not release the HOLD-key. Press also the ON/OFF/TARE-key and keep it
pressed
Keep the ON/OFF/TARE-key pressed, however release the HOLD-key
Press HOLD-key anew
Keep both keys pressed until in the display "tr" appears
Release both keys. The balance is situated in the menu.
Using the ON/OFF/TARE button you can select between the following functions:
tr (Zero tracking): ON/OFF
AF (Auto off): ON/OFF
rST (reset to factory settings): YES/NO
Press the HOLD button to confirm the selected function
In the display appears the current setting „ON" or „OFF" resp. „YES" or „NO". You
can select between „ON" or „OFF" and/or „YES" or „NO" with the
ON/OFF/TARE-button. Confirm your selection with the HOLD-key. After a short
period the balance returns automatically to weighing mode.