Télécharger Imprimer la page

OHAUS Adventurer Serie Guide De L'utilisateur page 13

Masquer les pouces Voir aussi pour Adventurer Serie:

Publicité

Les langues disponibles

Les langues disponibles

e)
Press Automatic Calibration to select On or Off.
Attention: Automatic Calibration must be set Off when the balance is used
for trade in Canada.
f)
Press Back to return to the previous menu or Exit to return to the Main
screen.
5.
Set the position of the security switch located at the back of the balance to the
locked position.
6.
Enter the Balance Setup menu and set the Approved Mode setting to On.
a)
Press Menu on the main screen.
b)
Press Balance Setup on the Main Menu screen.
c)
Press Approved Mode, and then press On.
d)
Press Back to return to the previous menu or Exit to return to the Main
screen.
6.2 Verification
A weights and measures official must perform the verification procedure. Contact
the local weights and measures office for more information.
6.3 Sealing
After the balance has been verified, the weights and measures official or authorized
agent will install an official seal to prevent undetected access to the legally
controlled settings. Before sealing the device, ensure that the security switch is in
the locked position and the Approved Mode setting in the Balance Setup menu has
been set to On.
If using a wire seal, pass the sealing wire through the holes in the security switch
and Bottom Housing as shown.
If using a paper seal, place the seal over the security switch and Bottom Housing as
shown.
Un-Locked
Locked with Wire Seal
Locked with Paper
Seal
EN-9

Publicité

loading