目次
1 Introduction
58 ページ
2 Specifications
58 ページ
第 1 章 Introduction
This document is an addendum to the CAx440EX Field Transmitter User Instructions. This document
supplies the safety precautions for installation of the CAx440EX field transmitter (LXV449.97.01010)
in hazardous locations.
Fully read the CAx440EX Field Transmitter User Instructions and GS1440, GS2440EX H
User Manual before the installation or operation of the sensor or field transmitter is started.
第 2 章 Specifications
Specifications are subject to change without notice. For the full specifications, refer to the CAx440EX
Field Transmitter User Instructions.
Specification
Safety/Hazardous location
Enclosure
Electrical ratings
Power/4–20 mA connector
(intrinsically safe outputs)
RS-232 connector
(intrinsically safe outputs)
Operating temperature
Storage temperature
Humidity
Altitude
North America certifications
(US/Canada)
UL standards
(US)
CSA standards
(Canada)
第 3 章 一般情報
いかなる場合も、製造元は、製品の不適切な使用またはマニュアルの指示に従わなかったことに起因
する損害について責任を負いません。製造元は、通知または義務なしに、随時本マニュアルおよび製
品において、その記載を変更する権利を有します。改訂版は、製造元の Web サイト上にあります。
58
日本語
3 一般情報
4 設置
Details
cETLus safety listed (5022889)
Polycarbonate, IP64/IP67
One of the options that follow:
• Battery: 3.6 V, 19 Ah 2 A maximum
• DC: 9–28 VDC, 1 A maximum
Uo: 29.2 V, Io: 85 mA, Po: 0.67 W, Lo: 4.9 mH, Co: 0.07 μF
Uo: 6.3 V, Io: 18.72 mA, Po: 0.30 W, Lo: 100 mH, Co: 14 μF
-20 to 60 °C (-4 to 140 °F)
-20 to 60 °C (-4 to 140 °F)
0 to 100% relative humidity
2000 m (6562 ft)
Class I, Division 2, Groups A-D, T4
Class I, Zone 2, Group IIC, T4
Equipment that provides intrinsically safe outputs:
[Ex ia] Class I, Division 1, Groups A-D
[AEx ia Ga] IIC and [Ex ia Ga] IIC
-20 °C ≤ Tamb ≤ +60 °C
Certificate: ETL23CA104847186X
Conforms to UL: 121201, 60079-0, 60079-11, 913, 61010-1
Certified to CSA C22.2: #213, #60079-0, #60079-11, #61010-1-12
58 ページ
61 ページ
S Sensor
2