FLUXUS H831
Tab. D.2 : Aperçu général des événements prédéfinis
ID
catégorie
message
2.07
F
Input: fuse broken
2.08
F
Input: power failure
2.09
F
Input: switched off
2.10
S
Input: temperature too high
2.11
S
Input: temperature too low
2.12
F
Input: measurement overflowed
2.13
S
Input: range exceeded
2.14
S
Input: below range
2.15
F
Input: communication error
2.16
F
Input: packet error
3.01
S
Output: range exceeded
3.02
S
Output: below range
3.03
F
Output: internal error
3.04
F
Output: communication error
3.05
F
Output: packet error
3.06
S
Output: temperature > 140 °C
3.07
S
Output: temperature > 100 °C
3.09
S
Output: max. current exceeded
3.10
S
Output: min. current not reached
3.11
S
Current output: supply voltage too low
3.12
S
Current output: supply voltage low
3.14
S
Current output: no voltage
3.15
S
Current output: internal voltage source error
UMFLUXUS_H831V1-4FR, 2023-10-01
D Auto-surveillance et diagnostic (NAMUR NE 107)
action
1. Replace input fuse
2. Replace input
Replace device
1. Reboot device
2. Replace input
Decrease temperature
Increase temperature
Check process input
Check input range
Check input range
1. Reboot device
2. Replace input
1. Reboot device
2. Replace input
Check output range
Check output range
1. Reboot device
2. Replace output
1. Reboot device
2. Replace output
1. Reboot device
2. Replace output
Decrease voltage
Decrease voltage
1. Passive output: decrease voltage
2. Active output: check if more than one active participant
is connected to the current output
3. Replace output
1. Passive output: increase voltage
2. Active output: check connection cable
3. Replace output
1. Increase external voltage
2. Shorten connection cable
1. Increase external voltage
2. Shorten connection cable
1. Connect external voltage supply
2. Check connection cable
1. Check connection cable
2. Reboot device
3. Replace output