Télécharger Imprimer la page

Siemens SITRANS FUH1010 Instructions De Service page 43

Débitmètres à ultrasons ip65 nema 4x & ip66 nema 7
Masquer les pouces Voir aussi pour SITRANS FUH1010:

Publicité

Les langues disponibles

Les langues disponibles

3. Wrap a Mounting Strap around the pipe. Make sure to position it so there is easy access
4. At the mounting location, place the Mounting Frame/Spacer Bar Assembly on the pipe so
5. Engage the end of the Mounting Strap with the Mounting Strap Adjusting Screw.
6. Slide strap under the spring clip of one of the Mounting Frames.
7. Tighten the Mounting Strap Screw enough to take up all of the slack, but not enough to
8. Rotate the assembly on the pipe to the final conditioned location, ensuring that it is
9. Tighten the mounting straps to seat the assembly firmly on the pipe. Do not over tighten.
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
to the Mounting Strap Adjusting Screw.
that it rests on the top of the pipe.
prevent rotation of the assembly.
straight along the pipe axis. (Refer to the sensor orientation diagram)
Repeat procedure for the other Mounting Frame
Connecting
3.5 Sensor Installation
.
41

Publicité

loading