your unit is connected to both AC and 12V DC power sources, it will operate on AC power as long as AC power is present
and automatically switch to 12V DC power if AC power is lost. When AC power is restored, it will return to the AC source.
You do not need to reset or adjust any settings when the power source switches, but remember that your heater will not
operate on 12V DC.
SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with
supplemental oxygen. Do not use this device near hot objects, volatile substances or sources of open flames.
Always turn on the IntelliPAP before turning on the oxygen source. Turn off the oxygen source before turning off the
IntelliPAP.
This device is not suitable for use in the presence of a flammable anesthetic mixture.
At a fixed flow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure
setting, patient breathing pattern, mask selection and leak rate. This warning applies to most types of CPAP devices.
Never allow the oxygen source to run continuously while connected to the IntelliPAP if the device is not in use. If the
IntelliPAP is not being used, turn off the oxygen flow.
CAUTION
Oxygen is a prescribed gas and should only be administered under the supervision of a physician.
The flow setting of the oxygen source must be specified by a physician.
Supplemental oxygen is sometimes prescribed in addition to a CPAP for treatment of Obstructive Sleep Apnea. If your
physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1.
Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the IntelliPAP. Connect one end of the air supply tubing to the open end of the adapter and the other end of the
air supply tubing to the mask.
2.
If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.
ACCESSORIES/REPLACEMENT ITEMS
Replacement Items
Air-inlet filter (4/pk)
Fine particle filter (4/pk)
Air supply tubing (6')
Air supply plug
Humidifier Chamber
Power Items
AC power cord (USA)
AC power cord (UK)
AC power cord (EU)
MAINTENANCE
Expected Service Life:
CPAP and humidifier cradle = 5 years
Humidifier water chamber = 1 Year
Tubing and gross particle intake filter = 180 days
DANGER
Electric shock hazard – Do not attempt to open or remove the cabinet. There are no user-serviceable components
inside. If service is required, contact your equipment provider for instructions on obtaining service. Opening or
attempting to service your device will void the warranty.
WARNING
Disconnect supply before servicing.
A-DV57
DV51D-602
Heater connection cover
DV51D-603
Oxygen adapter
CPAPTUB6-90
Carrying case
DV51D-604
Heated Humidifier
DV5C
Chamber Sealing Gasket
DV51D-606
AC power cord (Australia)
DV51D-608
DC power cord
DC battery clamp-on adapter (requires
DV51D-607
DC Power Cord DV51D-619)
DV51D-605
7353D-601
DV51D-610
DV5HH
DV5C-614
DV51D-609
DV51D-619
DV51D-696
EN - 11