SyMboLS

Refer to the manual
Conforms to relevant
Australian standards.
Warning and Precautions
when using Sound-Level Calibrator, ambient noise level should > 10dB
below the sound level being measured.

UNPACkINg ANd INSPECtIoN

Your shipping carton should include:
1
SM-CAL1 sound generator
1
1/2" adaptor
1
9 volt battery
1
Manual
if any of the items are damaged or missing, return the complete package to
the place of purchase for an exchange.

INtRodUCtIoN

The SM-CAL1 generates a fixed sound level signals for calibration of sound
level meters. The Amprobe SM-CAL1 will work with any sound meter.

oPERAtIoN

1.
Turn on sound level meter and configure for 94 dB or 114 dB operation.
2.
plug the 1/2 adapter into the SM-CAL1 Sound level calibrator.
3.
insert any sound meter microphone into the Sound-Level Calibrator.
4.
Turn on power switch and set sound level to 94 dB or 114 dB.
5.
observe the sound level meter for proper operation.