Environmental Conditions
Temperature range: 5 to 40°C (41 to 104°F)
Operating
Humidity range: 15 to 85%RH (not condensed)
Conditions
Atmospheric pressure: 700 to 1060hPa
Temperature range: -20 to 60°C (-4 to 140°F)
Storage and
Humidity range: 10 to 95%RH (not condensed)
transportation
Atmospheric pressure: 500 to 1060hPa
Non-Invasive Blood Pressure (NIBP)
Measurement
Oscillometric
technology
Measurement
Dynamic Linear Deflation method
method
Pressure
0 to 300 mmHg
display range
Pressure display
Within ±3 mmHg
accuracy
NIBP
SYS 60 to 250 mmHg
measurement
DIA 40 to 200 mmHg
range
PULSE 40 to 200 /min
Maximum mean error within ±5 mmHg
NIBP accuracy
Maximum standard deviation within 8 mmHg
Pulse rate
Within ±5 % of reading
accuracy
NOTE:
• Specifications may be changed without prior notice.
• The clinical accuracy of brachial pressure is implemented in the blood
pressure mode in accordance with ISO 81060-2: 2013. In the clinical
validation study, K5 was used on 88 subjects for determination of
diastolic blood pressure.
• This product conforms to EN80601-2-30: 2010+A1:2015.
• This blood pressure monitor conforms to the European
standard EN 1060, Non-invasive Sphygmomanometers Part 1:
General Requirements and Part 3: Additional Requirements for
Electromechanical Blood Pressure Measuring Systems.
• Please report to the manufacturer and the competent authority of the
Member State in which you are established about any serious incident
that has occurred in relation to this device.