Non-Invasive Blood Pressure (NIBP) Measurement
Measurement in "Normal Mode"
1. Press the [START/STOP] button.
Blood pressure measurement is performed once.
2. The measurement results are displayed.
If a measurement value is outside the corresponding range below, the
value will flash.
SYS: 59 mmHg or less, or 251 mmHg or higher.
DIA: 39 mmHg or less, or 201 mmHg or higher.
PULSE: 39 bpm or less, or 201 bpm or higher.
■ Normal measurement
Manual inflation in "Normal Mode"
If the cuff is not sufficiently inflated, it can be inflated manually.
During inflation, hold down the [START/STOP] button to inflate
continuously.
"-" appears below the value to indicate that manual inflation is in
progress.
18
■ Measurement error / failure