Software operation
Delayed heating
Starting the rotor preheating
28 / 39
CAUTION
Danger of burns from hot surfaces.
The centrifuging chamber and various parts of the housing
heat up.
−
Do not touch the centrifuging chamber or the relevant
parts of the housing.
NOTICE
Damage to plastic buckets due to excessive temperature
−
Plastic buckets may only be used at temperatures up to
a maximum of 40 °C or 104 °F.
The temperature in the centrifuging chamber increases
during the centrifugation run even when the heating is
switched off.
The temperature increase (sample heating) is dependent on
the rotor used, the set speed and the runtime.
A temperature lower than this cannot be achieved because
the device has no cooling system.
Recommendation: use the device for centrifugation runs
with temperatures from 40 °C to 90 °C.
If required, the settings can be adjusted so that heating takes place with a
time delay after the centrifugation run has started.
1.
If required: Press the
and RCF indicator.
[SELECT] button to select the delay time 'T delay/min' and
2.
Press the
[Rotary knob] button.
set it with the
The delay time is adjustable from 0 to 99 minutes, in 1-minute incre-
ments.
"0" must be set if no delay time is desired.
[SELECT] button to select the 'PROG STO' parameter.
3.
Use the
[Rotary knob] to set the desired program location.
4.
Use the
[START/IMPULS] button.
5.
Press the
Settings are stored in the desired program location.
'*** ok ***' is displayed briefly.
The settings are always stored in program location # if the
IMPULS] button is pressed without the 'PROG STO' parameter
being selected.
The centrifuge is started.
[Preheating] button.
1.
Press the
The
[START/IMPULS] button lights up during the centrifugation
run.
[STOP] button to stop preheating.
2.
Press the
Ramp-down takes place with the selected brake level.
The brake level is displayed.
Rev.: 09 / 11.2023
[RCF] button repeatedly to toggle between RPM
[START/