Télécharger Imprimer la page

Hettich ROTOFIX 46 Mode D'emploi page 61

Masquer les pouces Voir aussi pour ROTOFIX 46:

Publicité

Les langues disponibles

Les langues disponibles

5.4
Switching the centrifuge on and off.
Switching the centrifuge on
Switching off the centrifuge
6
Operation
6.1
Opening and closing the lid
Opening the lid
AB4600en
NOTICE
Damage to the device due to condensation.
There is a risk of condensation forming on electrical com-
ponents when component surfaces are cold and the sur-
rounding air is warmer. The condensation that forms may
cause a short circuit and/or destroy electronics.
Warm the device up for at least 3 hours in a warm room
before connecting it to the mains.
or
Warm up for 30 minutes in a cold room.
Personnel:
Trained user
1.
A type B residual current circuit breaker must be used if the device
is additionally protected with a residual current circuit breaker in the
building installation.
When using a different type, the residual current circuit breaker may
either not switch off the unit if there is a fault on the unit, or it may
switch off the unit even though there is no fault on the unit.
2.
Check whether the mains voltage matches the specification on the
rating plate.
3.
Connect the device to a standard mains socket using the mains
cable.
Personnel:
Trained user
Set the mains switch to
�� The buttons flash, depending on the centrifuge type.
The following indicators appear one after the other, depending on
the centrifuge type:
the centrifuge model and program version
When the lid is closed:
When the lid is open: The last centrifugation data used.
The rotor is stationary.
Set the mains switch to
Personnel:
Trained user
The centrifuge is switched on.
The rotor is stationary.
[Rotary knob, lid] on the front panel to the left.
1.
Turn
2.
Open the lid.
Rev.: 09 / 11.2023
[I] .
'OPEN OEFFNEN' indicator
[0] .
Operation
19 / 39

Publicité

loading

Ce manuel est également adapté pour:

Rotofix 46 h46004600-014600-504600-51