Operation
6.6.2
Lid with screw cap
Fig. 16: Biosafety system
1
Lid
2
Bucket
Closing
Opening
6.7
Centrifugation
6.7.1
Centrifugation in continuous operation
6.7.2
Centrifugation with time preselection
24 / 39
1
2
1.
Place the lid (
2.
Turn the lid (
1.
Turn the lid (
2.
Remove the lid (
Personnel:
■
Trained user
1.
Set minutes and seconds at
program.
[START/IMPULS] button.
2.
Press the
The centrifugation run is started.
[START/IMPULSE] button lights up during the centrifugation
The
run.
The timing starts at
The rotor speed or the RCF value, the temperature in the centri-
fuging chamber (only for centrifuges with heating) and the elapsed
time are displayed during the centrifugation run.
[STOP/OPEN] button to cancel the centrifugation run.
3.
Press the
Ramp-down takes place with the set brake level. The brake level is
displayed.
An audible signal sounds when the rotor comes to a standstill.
'OPEN' 'OEFFNEN' is displayed.
Personnel:
■
Trained user
1.
Set centrifugation parameters or retrieve a program.
[START/IMPULS] button.
2.
Press the
The centrifugation run is started.
[START] button lights up during the centrifugation run.
The
The rotor speed or the RCF value, the temperature in the cen-
trifuging chamber (only for centrifuges with heating) and the
remaining time are displayed during the centrifugation run.
Rev.: 09 / 11.2023
1 ) centrally on the bucket ( 2 ).
1 ) clockwise until it is tightly closed.
1 ) anticlockwise until it is open.
1 ) from the bucket ( 2 ).
'¥' or retrieve a continuously running
'00:00' .
AB4600en