Operation
Closing the lid
6.2
Removing and installing the rotor
Removing the rotor with a
clamping nut
4
Fig. 15: Rotor installation and
removal
1
Groove
2
Motor shaft
3
Driver
4
Hole
Installing the rotor with a
clamping nut
6.3
Inserting and removing buckets
Inserting buckets
20 / 39
NOTICE
Damage to the device caused by the lid slamming.
−
−
Personnel:
■
Trained user
1.
Close the lid and press the front edge of the lid down gently.
[Rotary knob, lid] on the front panel to the right.
2.
Turn
Personnel:
■
Trained user
1.
Open the lid.
2.
Loosen the rotor clamping nut using the supplied spanner.
After passing the working point for lifting the rotor, the rotor
detaches from the cone of the motor shaft (
1
3.
Turn the clamping nut until the rotor can be lifted off the motor shaft.
4.
Remove the rotor.
2
3
Personnel:
■
Trained user
The lid is open.
1.
Clean the motor shaft (
2.
Lightly grease the motor shaft (
cleaning and disinfection' on page 31 .
3.
Place the rotor vertically on the motor shaft (
3 ) of the motor shaft must be in the groove ( 1 ) of the
The driver (
rotor. The orientation of the groove is marked on the rotor.
4.
Hand-tighten the rotor clamping nut using the supplied spanner.
5.
Check that the rotor is firmly seated.
NOTICE
Damage to the device due to imbalances caused by incor-
rect loading of the rotor.
−
Rev.: 09 / 11.2023
Close the lid slowly.
Do not slam the lid.
2 ) and rotor hole ( 4 ).
2 ), see ⮫ Chapter 8.2 'Instructions for
Load all swing-out rotor locations with the same
buckets.
2 ).
2 ).