5.2
Removing the transport lock
AB4600en
NOTICE
Damage to the device due to improper lifting.
−
Do not lift the centrifuge by the control panel or the
control panel holder.
Personnel:
■
Trained user
1.
If present: Remove the packaging tapes.
2.
Lift the box up and remove the padding.
3.
Remove the accessories and store them safely.
4.
Place the device on a stable and level surface.
Personnel:
■
Trained user
5
1
4
2
4
Fig. 14: Transport lock
1
Screws
2
Motor cover
3
Bellows
4
Transport lock
5
Screws for transport lock
1.
Open the lid.
2.
Remove 4 screws ( ).
3.
Remove the motor cover ( ).
4.
For ROTOFIX 46 H:
Remove the bellows ( ).
5.
Remove 3 screws ( ) and 3 transport locks ( ).
6.
Keep the screws and transport locks in a safe place.
7.
For ROTOFIX 46 H:
Insert the bellows ( ).
Push the bellows ( ) over the edge of the rotor vessel. Note the recess
for the cable.
8.
Turn over the motor cover ( ) and screw it in.
Rev.: 09 / 11.2023
Commissioning
3
17 / 39
1
2