REFMATE operating instructions
General settings
Use the
tab for various
setting options.
The display brightness can be
adjusted as required.
Quick function:
Press and hold
tapping
or
brightness is adjusted in
increments of 5%.
Display Dimming
The display brightness can be
automatically reduced to
indicated level to save battery
power. The dimming occurs
after 2 Min.
Setting options: 5 – 95% and
off
Auto Shut Off
The time to auto power off is
set under Auto Shut Off. This
function can be deactivated.
Setting options: 1–60 min and
off
Heat Pump Mode
If the heat pump mode is set
to automatic (auto), the
analogue and digital display
switches automatically when
the pressure on the low
pressure side is higher than
on the high pressure side.
This function can be switched
"Off" and to "Automatic".
English
while
. The
19