6.1.6
Menu
Campo Calibrazione con
pressione di referenza
Zeroing the pressure sensors
Press and hold
The
pressure can also be zeroed in
the menu with zeroing the
pressure and press
.
Pressure calibration
In the navigation, go with
to
and press
. Use the
tab to carry out the field
calibration of the pressure
sensors.
To start the calibration, enter
the PIN. The PIN is 0-0-0-9
The device must not be
pressurised at this time!
Press
when
activated
Connect the device to a
system with a defined
pressure.
Pressure range: 10–60 bar
The higher the calibration
pressure, the more accurate
the calibration!
Wait until the calibration
pressure corresponds to the
reference pressure. By clicking
the respective pressure
sensor is calibrated.
101