5.
Use the Column UP/DOWN button or switch option to adjust the height of the CEPH
unit to approximately match the height of the patient.
Align the Ear Rods to the patient's ears properly so that the head does not move
6.
during the operation. And align the Nasal Positioner to the patient's nasion by adjusting
its height.
7.
Align horizontally so the patient's Frankfurt Line is parallel with the floor.
8.
Direct the patient to swallow first before closing the mouth and to remain in his/her
current position until image acquisition is completed.
9.
Click the READY button. The x-ray exposure has not started yet.
10. Now go to 4.3 X-ray Exposure to start the exposure.
Smart Plus User Manual
After adjusting the height of the column, align the Ear Rods and Nasal
Positioner to the patient.
4. Acquiring CEPH Images (Optional)