4.2.5.1
TILT mode configuration.
Pressing the CONFIG [14] key
will access to the first page of the
configuration menu.
To change a given parameter turn the
rotary selector [9] until the field is activated (this will
appear shadowed) and then press. The value of the
parameter will be activated and may be changed by
turning the rotary selector. When the value you
want appears, press the rotary selector [9] to
validate it.
a) BAND
It permits to select between the FORWARD mode (5 to 863 MHz) and the
REVERSE mode (5 to 80 MHz). When selecting one of these modes, the
user can define the associated frequency.
b) PILOT 1
It defines a frequency belonging to the forward or the reverse band where the
first measurement of level will be done. This parameter, and the rest of pilots,
is defined digit by digit, by pressing repeatedly and turning the rotary selector.
c) PILOT 2
It defines the second pilot frequency belonging to the forward or reverse
band.
d) PILOT 3
It defines the third pilot frequency belonging to the forward or reverse band.
e) PILOT 4
It defines the fourth pilot frequency belonging to the forward or reverse
band.
To leave the configuration menu of the TILT mode, just press the key of the
operating mode you wish to access.
Page 40
Figure 49.- TILT configuration. (REVERSE mode).
USER'S MANUAL.
PROMAX-10
Figure 48.- TILT configuration.
(FORWARD mode).
IMPORTANT
SE
07/2011