8. Troubleshooting
Trouble
The
SpO2
and Pulse
Rate display
instable
The device
can not turn
on
T
h
i n d i c a t o r
light is off
suddenly
ENGLISH
Possible Reason
1. The finger is not placed
inside enough.
2. The finger is shaking
or the patient is moving.
1. The batteries are drained
or almost drained.
2. The batteries are not
inserted properly.
3.
The
malfunction
e
1. The device will power
off automatically when it
gets no signal for 8
seconds.
2. The batteries are
almost drained.
Solution
1. Place the finger
properly and try again.
2. Let the patient keep
calm.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the
local service center.
device's
1. Normal.
2. Change batteries.
22