TB82PH ATEX Product Instruction Manual
CALIBRATE MODE
Before an instrument can properly measure pH the sensor must be calibrated to the instrument.
calibrations must be performed during the time the sensor is in service. The frequency of re-calibration is application
dependent. This mode also defines calibration of the temperature input and fine tuning the 4-20 mA output for non-
fieldbus versions.
MEASURE
CALIBRATE
Caution.
The
Reset
Calibrate
calibration values; therefore, the process sensor
and temperature sensor will require calibration
after performing the Reset Calibrate procedure.
OUTPUT/HOLD MODE
MEASURE
CALIBRATE
OUT/HOLD
Output/Hold screen flow only valid for non-fieldbus versions.
18
pH Cal
Temperature Cal
Edit Calibration
Reset Cal
Output Cal
Back to pH Cal
State
will
reset
all
Hold Output
Rerange Output
Damping
Back to Hold Output
1 Point Calibration
Stable?
Enter New Value
2 Point Calibration
Change Temp
Low Value Buffer
Stable?
High Value Buffer
Stable?
PV Slope %
PV Offset mV
Enter Temp
Enter Slope
Enter Offset
Adjust 4 mA
Adjust 20 mA
Release Hold
Hold to XXX%
4 mA Point
20 mA Point
XX.X Seconds
ABB
Periodic update
O
C
TB82PH-ATEX-EN-B