CONFIGURE MODE
Before the instrument can be put into a process measurement loop it must be configured. The TB82PH Transmitter
operation (non-fieldbus versions) can be set in either a BASIC or ADVANCED Mode. BASIC mode is sufficient for 95% of
all measurements.
ADVANCED functions are described in Italic print in the 'alternatives' section of the default
configuration table and in the 'Configure Mode - Flow Chart for ADVANCED Mode' section. Figures 15 through 25 show
typical screen flows for a BASIC pH Configuration.
Default Configuration
Parameter
Mode
Analyzer
Analyzer input
Temperature Sensor
Temperature
Compensation Type
Output
Damping
Diagnostics
Safe Mode*
Spike*
Note: Alternatives in Italics are available in ADVANCED mode only
7.00
CONFIG
exit to
MEASURE
SELECT
Figure 15
7.00
Basic
exit to
MEASURE
SELECT
Figure 17
TB82PH-ATEX-EN-B
Default
BASIC
pH
pH Glass
None
Manual
4 mA = 0.00 pH
20 mA = 14.00 pH
00.5 seconds
Off
Fail low
0% Spike Magnitude
* Safe Mode and Spike functions not available on fieldbus versions
MEASURE
CALIBRATE
pH
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
MENU
MEASURE
CALIBRATE
pH
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
NEXT
TB82PH ATEX Product Instruction Manual
Alternative
ADVANCED
ORP, pIon,
Ion Concentration
Antimony,
Custom sensor
Reference Impedance
3K Balco
Pt100 RTD
Auto,
Auto Solution
Any 1 pH to 16 pH span,
any span above 100 mV from
–1999 mV to +1999 mV,
non-linear function generator,
Adjustable to 99.9 seconds
On
Fail high
0 to 100% magnitude
7.00
MODIFY
exit to
MEASURE
SELECT
NEXT
Figure 16
7.00
ANALZR
exit to
MEASURE
SELECT
NEXT
Figure 18
MEASURE
CALIBRATE
pH
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
MEASURE
CALIBRATE
pH
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
15