ResMed AutoSet CS 2 Mode D'emploi page 36

Masquer les pouces Voir aussi pour AutoSet CS 2:

Publicité

Les langues disponibles

Les langues disponibles

W
P
OWER FAIL ALARM
T
I
F A POWER FAIL ALARM SOUNDS
W
CAUTION
!
Remove the mask from your face if the power fails.
H
IGH PRESSURE ALARM
T
I
F A
1. The treatment will stop.
2. Turn power off.
3. Check that the air tube and sensor tubes are connected properly.
4. Turn power back on.
5. Remove mask and perform Air Circuit Learn function.
6. Try using the flow generator one more time.
7. If the high pressure alarm activates repeatedly, discontinue use and return to
The alarm system is designed to ignore coughing. However, if a cough is especially intense
N
OTE
and prolonged, it may trigger the high pressure alarm.
W
H
IGH LEAK ALARM
T
I
F A HIGH LEAK ALARM SOUNDS
32
Get user manuals:
See SafeManuals.com
:
ILL CLEAR WHEN
the low pressure condition is fixed
treatment is stopped by pressing the Start/Stop key.
:
RIGGERED BY
power failure
machine is disconnected or switched off while delivering treatment.
The flow generator stops delivering air pressure.
:
ILL STOP
when the Alarm Mute key is pressed, or
after 2 minutes, or
when power is restored.
:
RIGGERED BY
the mask pressure exceeds the set trigger level (25 cmH
milliseconds.
H
IGH PRESSURE ALARM SOUNDS
ResMed for servicing. If the alarm does not recur, then continue to use as normal.
:
ILL STOP WHEN
the A
CS2 is turned off.
S
UTO
ET
:
RIGGERED BY
high mask leak (greater than 30 L/min) for more than 20 seconds.
Adjust the mask to minimise leak.
:
:
:
O) for more than 700