Télécharger Imprimer la page

WIKA GDT-20 Mode D'emploi page 50

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Annex 1: Accuracy of the dew point
Dew point accuracy
Measured value deviation < ±2 K for span T
Measured value deviation < ±3 K for span T
40
30
20
10
0
-40
-30
-20
-10
-20
-30
-40
-50
-60
Obergrenze (Messabweichung < ± 2 K)
Upper limit (Measuring deviation < ±2 K)
Lower limit (Measuring deviation < ±2 K)
Lower limit (Measuring deviation < ±3 K)
Examples:
If the gas temperature is +30 °C and the dew point is -20 °C, an accuracy of < ±2 K is
achieved (since Tgas - Td < 60 K).
If the gas temperature is +40 °C and the dew point is -30 °C, the accuracy of < ±3 K
is achieved (since Tgas - Td < 80 K)
50
Genauigkeit Taupunkt
Messwertabweichung < ± 2 K bei Spanne t
Messwertabweichung < ± 3 K bei Spanne t
-10
0
10
Gastemperatur t
Gas temperature T
Untergrenze (Messabweichung < ± 2 K)
WIKA operating instructions transmitter, models GDT-20 and GDHT-20
- T
< 60 K
gas
d
- T
< 80 K
gas
d
- T
< 60 K
Medium
d
- T
< 80 K
Medium
d
20
30
40
50
[°C]
[°C]
Medium
gas
Untergrenze (Messabweichung < ± 3 K)
60
70
80

Publicité

loading

Ce manuel est également adapté pour:

Gdht-20