Aesculap Suhl GmbH
Battery-powered clipping machine Favorita
4.4 Safe handling: Blade
4.4.1 Adjusting the plate pressure
Note
The plate pressure can be adjusted to increase the
cutting power as required (see foldout page)!
Tighten knurled screw 15 so that upper cutting
plate 19 can be moved by hand, by applying
moderate, lateral pressure.
4.4.2 Adjust plate pressure with Torqui pre-
cision adjusting aid
Note
The manufacturer recommends using the Torqui preci-
sion adjusting aid for optimal adjustment of the plate
pressure.
Insert Torqui precision adjustment aid 20 into the
socket hex of the knurled screw 15 and turn clock-
wise until a click can be heard.
Note
Knurled screws of older shearing heads that do not have
a socket hex can be retrofitted with the screw set
GT141.
4.4.3 Sharpening the cutting plate
Note
For sharpening the cutting plate, please contact
Technical Service, see Chapter Technical Service.
8
/Libra
CL
4.4.4 Removing the blade
Risk of injury caused by sharp
cutting plates!
Apply proper care when handling
WARNING
the blade of battery-powered
clipping
Libra.
Loosen knurled screw 15 with leaf spring 16.
Push upper cutting plate 19 under pressure
piece 14 out of the blade.
This completes the disassembly of the blade.
4.4.5 Lubricating the blade
Disassemble the blade, see Chapter Removing the
blade.
Clean upper cutting plate 19 and lower cutting
plate 18 with a soft wipe or brush. Make certain
the gaps in serration 17 are cleaned properly.
Slightly lubricate upper cutting plate 19 and lower
cutting plate 18 with special lubricant for blades
GT604.
machine
Favorita
/