Auto-setup function
12
All manuals and user guides at all-guides.com
Fig 7.0
Fig 8.0
VPS10 Panel Scope
AUTO-SETUP FUNCTION
The auto-setup function is ideally suited for quick
measurements as no manual setup has to be made
and everything is automatic. Use the auto-setup
function when the screen no longer displays a signal
after the manual setup.
Autorange on (Fig. 7.0) :
Keeping the 'Up' or 'Down' button pressed for a long
time enables autorange (only in V/div or t/div mode!)
Time/div and Volt/div settings are displayed inverted
The time base and input sensitivity are automatically
set for optimal viewing of the input signal.
Auto-triggering is set for time base 5µs/div or slower .
Normal triggering is set for time base faster then 5µs/div.
The slowest possible time base is 5ms/div.
The fastest possible time base is 1µs/div
Y-position is set to the center location.
Autorange off (Fig. 8.0) :
When autorange is enabled (time/div and V/div are
highlighted), it will be disabled as soon as the 'Up' of
'Down' button is used. (only in V/div, t/div or Y-pos.
mode!)
Time/div and Volt/div settings are displayed in a
normal font (not highlighted).