7 CALIBRATION METHODS (continued)
CALIBRATION METHOD
ISO 19840
SSPC PA2
Swedish
Australian
8 BATCHING
1
To use the Batching memory facility, press the Batch softkey
2
Select "New Batch" or "Open Existing Batch" to add readings
3
Copy, Rename, Review batch data and Clear readings from or Delete a batch
4
Fixed Batch Size allows users to pre-define the number of readings to be stored in a batch.
Once all readings have been taken the gauge automatically opens a new batch with a link
to the original batch name. For Example NewBatch becomes NewBatch_1, NewBatch_2, etc.
4
Save each reading into memory or store the average of a pre-defined number of readings
using the Counted Average function
7
All manuals and user guides at all-guides.com
MODEL DESCRIPTION
S, T
Sets the calibration method to Zero Offset and sets the
Counted Average to 5 - in accordance with ISO19840
S, T
Sets the calibration method to Rough/2 Point and sets the
Counted Average to 3 - in accordance with SSPC PA2
S, T
Sets the calibration method to Rough/2 Point and sets the
Counted Average to 5 - in accordance with SS standards
S, T
Sets the calibration method to Zero Offset, and sets the
Counted Average to 5 - in accordance with AS standards