Procedure:
Call-up menu item P3 CALCalLiner, see chap. 6.9
Confirm by
, the password query „Pn" will be displayed.
Press
,
,
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press
When "Ld 1" is displayed, put the first adjustment weight
(1/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press
When "Ld 2" is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press
When "Ld 3" is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform.
Wait for stability display, then press
"PASS" will be shown briefly, followed by the weight of the
placed adjustment weight.
Remove adjustment weight, balance will return into weighing
mode automatically. An adjusting error or incorrect
adjustment weight will be indicated by the error message;
repeat adjustment procedure.
18
subsequently.
.
.
.
.
KIB-TM-BA_IA-e-1811