7.6 Weighing with tare
Deposit weighing vessel. After successful standstill control press the
Zero display and indicator
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated.
The weight of the weighing container will be displayed as a minus number after
removing the weighing container.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The limit
is reached when the taring range capacity (see type plate)is full.
To change between gross weight and net weight, press
To delete the tare value, remove load from weighing plate and press
7.6.1 Pre-Tare
It is possible to enter a pre-tare value beforehand.
Ensure that there are no objects on the weighing pan.
Press
and hold until 0.0 is displayed and the left digit is flashing
Use the arrow keys to enter the pre-tare value and confirm by pressing
The pre-tare value will be shown as a negative value.
KIB-TM-BA_IA-e-1811
appear.
NET
button.
.
.
21