How to carry out adjustment:
Confirm menu setting „noLin" by
Ensure that there are no objects on the weighing plate.
Wait for stability display, then press
The currently set adjustment weight will be displayed.
Either use adjustment weight displayed
or
change with the help of the navigation keys (see chap.2.1.1),
the active digit will be flashing.
Confirm value for adjustment weight by pressing
Carefully place adjustment weight in the centre of the
weighing plate. Wait for stability display, then press
"PASS" will be shown briefly, followed by the weight of the
placed adjustment weight.
Remove adjustment weight, balance will return into weighing
mode automatically. An adjusting error or incorrect
adjustment weight will be indicated by the error message;
repeat adjustment procedure.
6.10 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range. If linearity deviation is discovered during a monitoring of test resources, you
can improve this by means of linearization.
• In balances with a resolution of > 15 000 dividing steps carrying out a
linearisation is recommended.
• Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
• The test weights to be used must be adapted to the weighing scale's
specifications; see chapter "Monitoring of test equipment".
• Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
• After successful linearisation you will have to carry out calibration;
see chapter 3.4 "Monitoring of test equipment".