Function
DC current (A=)
(min. 0,3 A)
AC current (A~)
(45 ... 65 Hz / min. 0,2 A)
DC voltage (V=)
(min. 0,5 V)
AC voltage (V~)
(45 ... 65 Hz / min. 1,0 V)
Frequency (Hz) AC current
(Measurement via Induction
clamp)
(40-1000 Hz /
Minimum 2 A eff.)
Frequency (Hz) AC voltage
(Measurement via Input
terminal)
(40-1000 Hz /
Minimum 0,8 V eff.)
Resistance (Ω)
*) Accuracy is valid for one year from the last calibration.
Range Resolution
60 A
0,01 µA
600 A
0,1 A
60 A
0,01 µA
400 A
0,1 A
600 A
0,1 A
6 V
0,001 V
60 V
0,01 V
600 V
0,1 V
6 V
0,001 V
60 V
0,01 V
600 V
0,1 V
60 Hz
0,1 Hz
1000
1 Hz
Hz
60 Hz
0,1 Hz
1000
1 Hz
Hz
6 kΩ
0,001 kΩ
60 kΩ
0,01 kΩ
600 kΩ
0,1 kΩ
6 MΩ
0,001 MΩ
10 MΩ
0,01 MΩ
EN 10
Accuracy of the
value displayed in %
*)
±(3,0% + 10 digits)
±(2,5% + 8 digits)
±(3,0% + 10 digits
±(0,5% + 3 digits)
±(0,8% + 5 digits)
±(1,0% + 5 digits)
±(1,0% + 5 digits)
±(0,8% + 3 digits)