Tecno-gaz ONYX B 6.0 Mode D'emploi page 94

Stérilisateur à vapeur

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 97
Fig. B
EXPOSURE TIME
CYCLES
T4 (Minutes)
18
121°C
4
134°C
4
134°C Fast
4
134°C Flash
4
134°C Safety
20
134°C Prion
134°C Prion
20
Fast
134°C Helix /
3.5
Bowie&Dick
-
Vacuum
The timing of the preheating and fractionating vacuum can vary depending on the autoclave from 25 - 35 minutes, these will add up with the timing of the cycles
shown in the table. All sterilization cycles have 3 vacuum phases except the 134 ° flash cycle that has 2 vacuum cycles. 
DRYING TIME
LOAD TYPE
T5 (Minutes)
wrapped and
15
unwrapped
wrapped and
15
unwrapped
wrapped and
10
unwrapped
4
unwrapped
4
unwrapped
wrapped and
15
unwrapped
wrapped and
10
unwrapped
4
-
-
-
WORKING
MAXIMUM LOAD
PRESSURE
(Kg)
RANGE
(bar)
7
1.04 ÷ 1.30
7
2.02 ÷ 2.38
2
2.02 ÷ 2.38
2
2.02 ÷ 2.38
7
2.02 ÷ 2.38
7
2.02 ÷ 2.38
2
2.02 ÷ 2.38
-
2.02 ÷ 2.38
-
-0.9
VOLUME OF
WORKING
WATER USED BY
TEMPERATURE
THE CYCLE (l)
RANGE (°C)
121 ÷ 124
0,83
134 ÷ 137
0,83
134 ÷ 137
0,83
134 ÷ 137
0,638
134 ÷ 137
0,83
134 ÷ 137
0,83
134 ÷ 137
0,83
134 ÷ 137
0,83
-