How does Braun No touch + touch work?
The No touch + touch thermometer measures infrared energy radiated from the skin at the center .
This captured energy is collected through the lens and converted to an oral equivalent value .
The No touch + touch thermometer has been clinically tested and proven to be safe and accurate
when used in accordance with its operating instruction manual.
Age Precision® colour-coded display
Clinical research shows the definition of fever changes as newborns grow into children and children grow
into adults .1
Select the appropriate age setting, take a temperature, and the display will glow green, yellow, or red to
indicate whether the temperature is in a normal, fever, or high fever range, respectively .
The device is only intended for the measurement of human body temperature . The device also has a food/
bath mode . It is not intended to diagnose or monitor a serious medical condition or disease .
1 Herzog L . ., Phillips S . Addressing Concerns About Fever . Clinical Pediatrics . 2011; 50 (#5): 383-390 .
Temperature taking hints
It is important to know each individual's normal temperature when they are well . This is the only way to
accurately diagnose a fever . Take multiple readings when healthy to determine normal temperature .
A child's normal temperature can be as high as 37 . 7 °C (99 . 9 °F) or as low as 36 .1 °C (97 .0 °F) . Be sure to note this
unit reads 0 .5 °C (0 .9 °F) lower than a rectal digital measurement .
An individual must be inside for 30 minutes before taking a measurement.
Note: The thermometer and individual should be in the same ambient temperature
for at least 10 minutes.
ALWAYS hold the thermometer and the forehead steady when taking a reading . DO NOT move the
thermometer until you hear the final beep . DO NOT take a measurement while or immediately after
nursing a baby .
Individuals should not drink, eat, or be physically active before/while taking the measurement .
Remove hats and wait 10 minutes before taking a measurement .
Before taking a measurement, remove dirt or hair from the forehead area . Wait 10 minutes after cleaning
before taking measurement .
ALWAYS take the temperature exactly as directed . Temperature results may vary if positioned in
the wrong location .
For individuals measuring their own temperature, it is recommended to use the "touch" option instead
of "no touch . "
In the following situations it is recommended that three temperatures in the same location be taken and
the highest one taken as the reading:
Newborn infants in the first 100 days .
Children under three years of age with a compromised immune system and for whom the presence
or absence of fever is critical .
4
When the user is learning how to use the thermometer for the first time until he/she has familiarized
himself/herself with the instrument and obtains consistent readings .
General precautions:
Remove any sweat prior to measuring with a dry cloth .
Avoid any cooling or warming cloths on the forehead for at least 30 minutes prior to measurement .
DO NOT take temperature measurements over scar tissue, open sores, or abrasions .
Keep the measurement sensor and lens clean . Avoid directly touching the sensor or lens .
ALWAYS store the thermometer with the protective cap in place to prevent dirt and scratches from
damaging the device .
Make sure to remove the protective cap before taking a measurement and to put the cap back on after
using the thermometer .
How to use your Braun No touch + touch
Note: The thermometer and Individual should be in the same ambient temperature for at
least 10 minutes.
Remove protective cap
1
Power on
2
Press and release the power button once . Backlight will come
on and the start-up sequence begins .
NOTE: Make sure to remove the protective scanner cap
before taking a measurement .