ME10
4.2 Setting the Measuring Range
Measuring range 1 is get factory settings. Measuring range 2 can be activated
by switching DIP switch S24/4 or by external switching.
Measuring Range
MR 1 = "2 mV/V
(factory settings)
MR 2 = "0.2 mV/V
Any alteration to the measuring ranges between "0.2 mV/V and "4 mV/V
with U
= 5 V (U
B
B
by means of the DIP switches S26 and S27. For fine balancing see 3.3.
U
= 5V
B
Meas. range
(mV/V)
0,17
. . .
0,24
0,22
. . .
0,32
0,30
. . .
0,42
0,39
. . .
0,55
0,52
. . .
0,73
0,68
. . .
0,97
0,90
. . .
1,28
1,18
. . .
1,69
1,56
. . .
2,23
2,07
. . .
2,94
2,73
. . .
3,88
3,61
. . .
5,13
o = open, x = closed
A0762−4.0 en/de/fr
= 2.5 V, resp. U
B
MB1:S26 ; MB2:S27
1
2
3
x
o
o
x
o
o
o
x
o
o
x
o
o
o
x
o
o
x
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
S 24/4
= 10 V, see Technical Data) can be made
4
5
6
7
o
o
o
x
o
o
o
o
o
o
o
x
o
o
o
o
o
o
o
x
o
o
o
o
x
o
o
x
x
o
o
o
o
x
o
x
o
x
o
o
o
o
x
x
o
o
x
o
Externally
−
from MR1 to MR2;
connect 25c and 15c
factory settings
8
o
MB 2 "0,2 mV/V
x
o
x
o
x
o
x
o
MB 1 "2 mV/V
x
o
x
15
HBM