14
4
Individual Settings
Various factory settings settings on the measuring amplifier card can be al-
tered. Adaptations, which are of no problem, can be carried out by means of
the DIP switch on the board or also by externally wired switching elements. If
requested with the order, they can be carried out by the manufacturer.
4.1 Operational Setting
S21
Switch
The transducer signal is switched off in the "Zero" function, so that the "Zero
Value" of the amplifier is applied to the output. This function enables amplifiers
to be replaced in measuring chain without the need for rebalancing the bridge
with the unloaded transducer. All that is required is that the amplifier balance
values are noted and then this value is set, if necessary, with P23, S22 and
S23.
HBM
Zero
or connect pin 21a to pin 15c
or connect pin 25a to pin 15c
Measure
Calibrate
ME10
A0762−4.0 en/de/fr