3.3 MR Fine Adjustment
The fine adjustment of the measuring range is made seperately for both
ranges by means of the screwdriver potentiometers P21 (measuring range 1)
and P22(measuring range 2). The potentiometers in the front panel permit
corrections of approx. 35 % (related to the set measuring range). As far as
measuring range 1 is concerned, the fine adjustment is also possible via an
external potentiometer (for this see chapter 4.3).
NOTE
P22 must not be altered in order not to detune the manufacturer's setting
(MR2 "0.2 mV/V) (see chapter 4.5).