Putting into Operation
If the factory settings, such as the coarse measuring range, are retained the
operation is limited to the following points:
3.1 Select Measuring Range
With the factory setting measuring range 1 ("2 mV/V) is always switched on.
Measuring range 2 (calibration at "0.2 mV/V) can be selected either internal-
ly via DIP switch 24/4 or externally by closing a single-pole change-over
switch to pin 25c against operating voltage zero (see chapter 4.2).
3.2 Bridge Zero Balancing
Carry out bridge zero balancing by means of the screwdriver potentiometers
P23 (fine) and switches S22 (coarse) and S23 (polarity) (see also chapter
4.4).
HBM
ME10
A0762−4.0 en/de/fr